    A resource for the development of methodologies for lung nodule size estimation: database of thoracic CT scans of an anthropomorphic phantom

    MIDAS/ISP: Lung Cancer Imaging/Article #120091

    Title:A resource for the development of methodologies for lung nodule size estimation: database of thoracic CT scans of an anthropomorphic phantom
    Authors:Marios A. Gavrielides1, Lisa M. Kinnard2, Kyle J. Myers1, Jennifer Peregoy1, William F. Pritchard1, Rongping Zeng1, Juan Esparza1, John Karanian1, Nicholas Petrick1
    Institutions: 1U.S. Food and Drug Administration, 2Congressionally Directed Medical Research Program
    Publisher:Optical Society of America
    Publication date: Received 2009-Nov-18
    PDF Article: 120091.pdf
    Abstract:A number of interrelated factors can affect the precision and accuracy of lung nodule size estimation. To quantify the effect of these factors, we have been conducting phantom CT studies using an anthropomorphic thoracic phantom containing a vasculature insert to which synthetic nodules were inserted or attached. Ten repeat scans were acquired on different multi-detector scanners, using several sets of acquisition and reconstruction protocols and various nodule characteristics (size, shape, density, location). This study design enables both bias and variance analysis for the task on nodule size estimation. The resulting database provides a publicly available resource to facilitate the assessment of lung nodule size estimation methodologies.
    Complete Data Collection & Article:article_860_full.zip (43Mb)
    Compressed Data Collection: article_860_preview.zip (3Mb)
    Data Collection:
    View1Fig. 4
    Fig. 4
    Fig. 4
    Fig. 4
    Case001
    Case002
    Case003
    Case004
    FDA phantom CT-Example1
    FDA phantom CT-Example1
    FDA phantom CT-Example1
    FDA phantom CT-Example1
    View2Fig. 5
    Fig. 5
    Fig. 5
    Fig. 5
    Case005
    Case006
    Case007
    Case008
    FDA phantom CT-Example2
    FDA phantom CT-Example2
    FDA phantom CT-Example2
    FDA phantom CT-Example2
    View3Fig. 6Case009FDA phantom CT-Example3
    OCIS Code: 110.6880 , 170.7440 

    © 2025 Optical Society of America



    Show bitstreams Hide bitstreams


      Please wait...
      • Navigation

        loading
        /midaspre/
      • Browse

        • Communities & Collections
        • Image Gallery
        • Titles
        • Authors
        • Date
        • Keywords
      • More information

        • About ISP
        • Get OSA ISP Software
        • Terms of Use
        • About MIDAS
      • OSA Links

        • Optics InfoBase
        • OSA - The Optical Society

      MIDAS 2.5.0 (module osa 2.3.0) by Kitware Inc © 2009
      Kitware Logo